Muestra métricas de impacto externas asociadas a la publicación. Para mayor detalle:
| Indexado |
|
||||
| DOI | 10.2147/OPTH.S105065 | ||||
| Año | 2016 | ||||
| Tipo | artículo de investigación |
Citas Totales
Autores Afiliación Chile
Instituciones Chile
% Participación
Internacional
Autores
Afiliación Extranjera
Instituciones
Extranjeras
Conclusion: In all experiments for each tip design, increasing ultrasound power was associated with greater increases in tip temperature. Tip temperatures increased with applied load, but marked temperature increases during occlusion were observed only with the bent-mini tip. The balanced tip produced minimal thermal peaks in all tests.
| Ord. | Autor | Género | Institución - País |
|---|---|---|---|
| 1 | Zacharias, Jaime | Hombre |
Pasteur Ophthalm Clin - Chile
Phacodynamics Laboratory - Chile |
| Agradecimiento |
|---|
| This study was sponsored by Alcon Laboratories, Inc. (Fort Worth, TX, USA). The author thanks Ramon Dimalanta, PhD, of Alcon Research, Ltd (Irvine, CA, USA), who aided in the completion of this study. Scientific writing assistance was provided by Heather D Starkey, PhD, of Complete Healthcare Communications, Inc. (Chadds Ford, PA, USA) and was funded by Alcon Laboratories, Inc. Alcon participated in revising the manuscript and supported the decision to submit the manuscript for publication. |