Entrance surface air kerma (ESAK) values and image quality parameters were measured and compared for two biplane angiography x-ray systems dedicated to paediatric interventional cardiology, one equipped with image intensifiers (II) and the other one with dynamic flat detectors (FDs). Polymethyl methacrylate phantoms of different thicknesses, ranging from 8 to 16 cm, and a Leeds TOR 18-FG test object were used. The parameters of the image quality evaluated were noise, signal-difference-to-noise ratio (SdNR), high contrast spatial resolution (HCSR) and three figures of merit combining entrance doses and signal-to-noise ratios or HCSR. The comparisons showed a better behaviour of the II-based system in the low contrast region over the whole interval of thicknesses. The FD-based system showed a better performance in HCSR. The FD system evaluated would need around two times more dose than the II system evaluated to reach a given value of SdNR; moreover, a better spatial resolution was measured (and perceived in conventional monitors) for the system equipped with flat detectors. According to the results of this paper, the use of dynamic FD systems does not lead to an automatic reduction in ESAK or to an automatic improvement in image quality by comparison with II systems. Any improvement also depends on the setting of the x-ray systems and it should still be possible to refine these settings for some of the dynamic FDs used in paediatric cardiology.
| Ord. | Autor | Género | Institución - País |
|---|---|---|---|
| 1 | Vano, E. | Hombre |
UNIV COMPLUTENSE - España
San Carlos Univ Hosp - España Universidad Complutense de Madrid - España |
| 2 | UBEDA-DE LA CERDA, CARLOS ARTURO | Hombre |
Universidad de Tarapacá - Chile
|
| 3 | Martinez, Luis C. | Hombre |
12 Octubre Univ Hosp - España
Hospital Universitario 12 de Octubre - España |
| 4 | LEYTON-LEGUES, FERNANDO ALBERTO | Hombre |
Inst Publ Hlth Chile - Chile
Institute of Public Health of Chile - Chile |
| 5 | MIRANDA-GONZALEZ, PATRICIA | Mujer |
Hospital Luis Calvo Mackenna - Chile
Hospital Dr. Luis Calvo Mackenna Hospital - Chile Hosp Ninos Luis Calvo Mackenna - Chile |
| Fuente |
|---|
| Ministry of Science and Innovation |
| Direction of Research at Tarapaca University |
| Agradecimiento |
|---|
| The authors thank the Mr Bernard Geiger (Siemens AG, Healthcare Sector, D-91301 Forchheim, Germany) for his fruitful discussions and comments on this paper. EV acknowledges the support of the Spanish grant SAF2009-10485 (Ministry of Science and Innovation). CU acknowledges the support of the Direction of Research at Tarapaca University through senior research project no 7713-10. |